Newsroom
VHP pass box for pharmaceutical manufacturing: how to choose the right configuration
A VHP pass box transfers heat-sensitive materials from a lower-grade area into a cleanroom, for example from EU GMP grade C to grade B, and bio-decontaminates their exposed surfaces with vaporized hydrogen peroxide at low temperature. The configuration of this pass-through chamber depends on the load, the chamber volume, the pressure regime of the cycle and the interface with the cleanroom wall. The sections below describe how each of these elements shapes the choice and how they are resolved on LAST Technology DPB pass boxes.
The VHP process principle, the cycle phases and the differences from aerosolized hydrogen peroxide and gas plasma are covered in the article VHP decontamination in pharmaceutical manufacturing.
What a VHP pass box does in a cleanroom transfer
A VHP pass box is used when materials that tolerate neither saturated steam nor dry heat have to enter grade A or B areas and cannot be sterilized at the point of transfer. The typical case is material sterilized in sealed packaging, whose outer surface is treated before it enters the cleanroom.
Its working principle is an automatic bio-decontamination cycle that runs between the closing of the loading door and the opening of the unloading door: hydrogen peroxide vapor is distributed over the load, held for the exposure phase and then removed before the material is released on the clean side.
Pass boxes used in pharmaceutical cleanrooms differ in what happens to the load during the transfer:
| Pass box type | Effect on the load surfaces |
| Static, no air supply | No treatment during the transfer |
| Dynamic, HEPA-filtered air | Chamber flushed with filtered air, no sporicidal action on the load |
| UV lamps | Action limited to the surfaces directly reached by the light |
| VHP | Automatic bio-decontamination cycle, with validated microbial reduction on the exposed surfaces |
LAST Technology DPB pass boxes belong to the fourth category: they work with air and hydrogen peroxide between 20 and 30 °C (68 to 86 °F), with a cycle customized to the load being treated.
Where EU GMP Annex 1 places the VHP pass box
EU GMP Annex 1 (2022 revision) treats the transfer of materials into clean areas as one of the main potential sources of contamination (paragraph 4.10) and addresses it in four points that apply directly to a VHP pass-through chamber.
- Unidirectional transfer (4.11). Materials entering grade A and B areas follow a unidirectional process. Where possible they are sterilized in double-ended sterilizers sealed into the wall; where that is not possible, a validated procedure with the same objective is required, such as a transfer disinfection process.
- Approved materials and protective design (4.12). Only materials on an approved list, assessed during validation of the transfer, enter grade A and B. Pass-through hatches are designed to protect the higher-grade environment, for example through active flushing with filtered air.
- Door control (4.13). The doors of pass-through hatches and airlocks are not opened at the same time; transfers into grade A and B require an interlock system and, where needed to keep the areas separated, a delay between the closing of one door and the opening of the other.
- Sealed sterilized items (8.47). Materials sterilized in sealed packaging and transferred into grade A through an airlock or pass-through undergo disinfection of the outer packaging, with its effectiveness demonstrated.
The load as the starting point of the configuration
In VHP bio-decontamination the result applies to the surfaces the vapor reaches. Material, packaging and arrangement on the carts therefore determine chamber volume, pressure regime and aeration time.
| Load type | What it involves |
| Plastic syringe trays, electronic components, pre-sterilized material | Low-temperature treatment of the outer surfaces |
| Closed containers holding liquids, such as disinfectants | The cycle treats the outer surface of the container; the liquid inside is not treated |
| Multi-layer packaged materials | The cycle acts on the outer layer; the content remains protected by the upstream sterilization |
| Plastics and absorbent materials | Longer aeration, because absorbed peroxide is released slowly |
| Items with cavities or stacked items | Hard-to-reach surfaces, included in the claim only if demonstrated on the representative load |
| Liquids to be treated in their volume | Outside the scope of VHP: for liquids and loads that tolerate heat and moisture, saturated steam sterilization remains the reference |
A new load, or a change in the arrangement on the carts, can require new cycle development and a new qualification. Saturated steam sterilization is described in the article on saturated steam cGMP sterilization with RSA autoclaves.
Chamber volume and load handling
On DPB pass boxes the chamber volume and internal geometry follow the load to be transferred, not a standard size. The construction elements that affect the configuration are:
- chamber and doors with square or rectangular cross-section, in AISI 316L stainless steel
- rounded internal edges and mechanically polished surfaces, then degreased, pickled and passivated, with a surface finish of ≤ 0.35 µm
- height-adjustable guides on the chamber floor for loading and unloading the carts
- tubular base frame that distributes the weight evenly on the floor
- installation on the floor or in a pit for larger volumes
Chamber volume is one of the factors that determine the amount of hydrogen peroxide and the cycle duration, together with generator capacity, vapor distribution, temperature and load materials.
Atmospheric or vacuum VHP cycle
A VHP cycle runs either close to atmospheric pressure or under vacuum. In both cases the agent is hydrogen peroxide vapor; vacuum removes the air before injection and changes the way the vapor reaches the load.
| Factor | Atmospheric pressure | Vacuum |
| Load | Accessible outer surfaces | Geometries that trap air; the result still has to be demonstrated on the representative load |
| Vapor distribution | Relies on internal ventilation | Chamber saturation without air diluting the vapor |
| Chamber | Not subject to negative pressure | Designed to withstand vacuum, with effects on feasible dimensions and cost |
| Sensitive loads | No pressure change | Some loads and packaging do not tolerate vacuum |
In the LAST Technology range the vacuum VHP cycle is available on RSA HP autoclaves, which are not a DPB variant: they combine, in the same chamber, saturated steam sterilization of porous and non-porous loads and liquids with VHP bio-decontamination of non-porous materials. Their features are described on the saturated steam autoclave + bio-decontamination with VHP cGMP page.
Cleanroom interface, doors and interlocks
The pass box connects two areas with different air cleanliness grades: the loading side is in the less clean area, the unloading side in the cleanroom. On DPB pass boxes the unloading door is designed for connection to the cleanroom wall through a Bio-seal interface.
The interlock prevents both doors from opening at the same time; the direction of material flow follows from the position of the two sides and from site procedures.
On DPB pass boxes the doors are interlocked: while one door is open, the other stays locked. The unloading door is also tied to the cycle outcome. If the cycle is interrupted or fails, only the loading door can be opened, so the material does not reach the sterile area. Both doors stay locked until the hydrogen peroxide concentration in the chamber falls below the safety level of 1 ppm. Doors can be hinged or sliding.
Aeration and release criterion
On DPB pass boxes the hydrogen peroxide generator is integrated into the machine. An internal ventilation system distributes the vapor over the load and, at the end of exposure, a catalytic degassing system breaks the residual peroxide down into water and oxygen.
The peroxide concentration in the chamber is monitored by two separate sensors: the first measures the level reached during decontamination, the second, a low-range sensor, checks the safety level that enables door opening. Aeration time depends mainly on the load: plastics and absorbent materials release hydrogen peroxide more slowly than stainless steel.
VHP pass box qualification and validation
On DPB pass boxes the predefined cycle programs are developed by the LAST Technology Automation department according to the applicable standards and the product to be decontaminated; custom programs are created by the customer. Two 1.5″ tri-clamp ferrules are welded to the chamber for the insertion of validation instruments.
The microbial reduction a bio-decontamination pass box can demonstrate derives from the validated cycle, the load configuration and the biological indicators used, and is established during cycle development together with the customer’s quality function. The biological indicator validation method is described in the VHP article linked at the beginning.
DPB pass box design data
The configuration of a DPB starts from a defined set of design data. The table shows what each one determines and how it is resolved on the DPB series.
| Design data | What it determines | Solution on DPB |
| Load materials, packaging, geometry and weight | Surfaces to be treated, pressure regime, aeration time | Cycle developed on the reference load defined with the customer |
| Dimensions of load and carts | Usable chamber and door dimensions | Square or rectangular chamber sized on the load, height-adjustable guides |
| Transfers per shift | Chamber volume and number of cycles required | Assessed with the customer at design stage |
| Available space and floor level | Installation type | On the floor or in a pit for larger volumes, on a tubular base frame |
| Grades of the two areas and wall type | Interface, doors and interlocks | Unloading door with Bio-seal interface toward the cleanroom |
| Qualification and data management requirements | Instrumentation, programs, cycle records | Predefined LAST Technology programs and customer-created custom programs; 1.5″ tri-clamp ferrules for validation |
The DPB series is a VHP process: no configuration includes a plasma phase.
Conclusion
LAST Technology DPB pass boxes bio-decontaminate the surfaces of heat-sensitive materials bound for the cleanroom with vaporized hydrogen peroxide, between 20 and 30 °C, with a chamber sized on the load and an unloading door connected to the wall through a Bio-seal interface. The construction specifications of the series are on the cGMP bio-decontamination pass boxes page; the technical office can be reached at info@lasttechnology.it.
References
- European Commission, EU GMP Annex 1, Manufacture of Sterile Medicinal Products, 2022 revision, paragraphs 4.10-4.13 and 8.47
FAQ - Domande frequenti
A VHP pass box is a pass-through chamber installed in the cleanroom wall that bio-decontaminates the exposed surfaces of heat-sensitive materials with vaporized hydrogen peroxide before they enter the clean area. It combines the transfer function of a pass box with an automatic, validated bio-decontamination cycle.
Related posts
mail
Subscribe to our newsletter
Stay up to date with the latest news from LAST Technology